Matrix logo
MDM-8200 Series High-Precision Digital Multimeter

MDM-8200Series High-Precision Digital Multimeter

  • 6½ digital display readings
  • 4.3-inch 64K color TFT LCD screen
  • Multiple display forms
  • Dual display measurement function
  • 0.1uV high resolution
  • Various measurement functions

Want a custom configuration?

This multimeter has outstanding practical and safety features: front and rear dual input ports facilitate wiring of test systems, supporting SCPI remote control and mainstream command sets, with standard upper computer software. Interfaces include USB, RS232/485, LAN, etc., enabling flexible and efficient operation. It has functions such as data statistics and mathematical operations, with 0.1uV high resolution to detect subtle signals. The safety standard reaches CAT II 300V, with built-in comprehensive protection mechanisms, balancing high-precision measurement and operational safety, easily meeting professional measurement needs in R&D, quality inspection, and other fields.

Series SpecMDM-82656 1/2 Reading Res, 30,000 readings/s max, I/O: USB/RS232/485/LAN/GPIB, 75 PPM base
MDM-82555 1/2 Reading Res, 30,000 readings/s max, I/O: USB/RS232/485/LAN/GPIB, 150 PPM base
DC VoltageMDM-8265 Accuracy (100mV to 1000V)Best: 0.0035+0.0005 (10V range), Min Res: 0.1μV
MDM-8255 Accuracy (100mV to 1000V)Base: 0.015+0.005, Min Res: 1μV
ResistanceMDM-8265 (100Ω to 100MΩ)Best: 0.010+0.001 (10kΩ), Min Res: 100μΩ
MDM-8255 (100Ω to 100MΩ)Best: 0.050+0.005 (10kΩ), Min Res: 1mΩ
DC CurrentMDM-8265 (100μA to 10A)Best: 0.050+0.005 (10mA), Min Res: 1nA
MDM-8255 (100μA to 10A)Best: 0.050+0.007 (1mA), Min Res: 10nA
AC VoltageMDM-8265 (Frequency 5Hz to 300kHz)Best: 0.06+0.03 (10Hz–20kHz) at 100mV-750V
MDM-8255 (Frequency 20Hz to 100kHz)Best: 0.20+0.10 (45Hz–10kHz) at 100mV-750V
AC CurrentFrequency: 100μA-10AMDM-8265 (3Hz–10kHz, Best: 0.10+0.04) | MDM-8255 (20Hz–1kHz, Best: 0.50+0.10)
Other Meas.Capacitance (1nF to 100uF)MDM-8265: 0.050+0.50 | MDM-8255: 1+0.5
Temperature (PT100, 5K thermistor)Probe accuracy + 0.05℃ (PT100) or + 0.1℃ (Thermistor)