Intelligent and Efficient Testing Solution This series of testers features a 4.3-inch TFT color LCD screen with an intuitive and easy-to-understand interface. It supports programmable storage of 105 test files, each capable of setting 25 consecutive test steps to meet the automated testing needs of complex scenarios. Equipped with rich interfaces such as RS232C, USB, and HANDLER, it enables computer remote control, data export, and quick import of test plans, seamlessly connecting production line and laboratory scenarios. Parameters such as voltage rise/fall time and test delay can be precisely adjusted (0.1s-999.9s). Combined with the optional multi-channel module, continuous testing of multiple measurement points can be completed without repeated wiring, significantly improving testing efficiency. With a compact design (215×143×405mm) and a weight of only about 12kg, it balances portability and stability, making it an ideal testing partner for electronic manufacturing, quality inspection, R&D verification, and other scenarios.
| Function Description | MST-8101: AC | MST-8103: AC/DC/IR | |
| AC Withstand Voltage | Voltage range | 0.050kV—5.000kV |
| Voltage waveform | Sine wave | |
| Distortion | <3% | |
| Working frequency | 50/60Hz optional | |
| Frequency accuracy | ±1% | |
| Output Power | 100VA (20mA) | |
| Voltage regulation rate | ±(1.0%+50V) (rated power) | |
| DC Withstand Voltage (MST-8103) | Voltage range | 0.050 kV—6.00kV |
| Signal source frequency | 600Hz | |
| Output Power | 50VA (10mA) | |
| Voltage regulation rate | ±(1.0%+100V) (rated power) | |
| General Voltage Params | Voltage resolution | 1V |
| Voltage test accuracy | ±2% | |
| Voltage generation method | DDS signal source plus class AB power amplifier | |
| AC Current Measurement | Current range | 0.001mA - 20.00 mA |
| Short circuit current (momentary) | >40 mA | |
| Current resolution | 0.001 mA | |
| Current accuracy | ±(2% of reading + 2 words) | |
| Actual current | OFF - 0.001 mA - 20 mA | |
| DC Current Measurement (MST-8103) | Current range | 0.1uA - 10.00mA |
| Current accuracy | ±(2% reading + 2 digits) | |
| Discharge function | Automatic discharge after the test (DCW) | |
| Insulation Resistance (MST-8103) | Output voltage | 0.050V - 1.000kV |
| Voltage resolution | 1V | |
| Voltage test accuracy | ±2% | |
| Maximum output current | 10mA | |
| Maximum output power | 10VA (1000V/10mA) | |
| Output instantaneous short circuit current | >20mA | |
| Load Regulation | ≤1% (rated power) | |
| Ripple (1kV) | ≤3% (1kV, no load) | |
| Discharge function | Automatic discharge after the test | |
| Resistance measurement range | 0.1MΩ - 10GΩ | |
| Resistance measurement accuracy | Voltage < 500V: 0.2MΩ~1GΩ ±10%, 1GΩ~10GΩ ±20%; Voltage > 500V: 0.2MΩ~1GΩ ±3%, 1GΩ~10GΩ ±7% | |
| Arc Detection | Measuring range AC/DC | AC: 1mA-20mA; AC/DC: 1mA-20mA |
| Comparators | Discrimination method | Window comparator mode |
| Capping current I on | AC: 0.001mA-20mA, DC: 0.1uA-10mA | |
| Current upper limit setting I under | AC: 0.001mA-20mA, DC: 0.1uA-10mA | |
| Resistance upper limit setting | OFF-0.2MΩ-10GΩ | |
| Resistance lower limit setting | 0.2MΩ-10GΩ | |
| Parameter Setting | Voltage rise time | 0.1s - 999.9s |
| Voltage drop time | 0s - 999.9s (only after the withstand voltage PASS) | |
| Voltage waiting time | 0.3s - 999.9s (DC withstand voltage only) | |
| Test time setting | 0.3s - 999.9s (when TIMER ON) | |
| Time accuracy | ±(0.2% setting value ± 0.1s) | |
| Other | Protocol | SCPI, Modbus |
| Storage | 105 test files can be programmed, and 25 test steps can be set for each file | |
| Interface | HANDLER, SINGAL, RS232C, RS485 (optional) | |
| Size (W*H*D) | 215 * 143 * 405 mm | |
| Weight | 12 kg |