This oscilloscope has outstanding practical features: it supports dual-channel simultaneous display (alternate/intermittent mode switching), X-Y operation, and TV synchronization function, enabling observation of TV field/line signals or two unrelated signals. The ALT trigger function ensures stable display of signals with different frequencies. The input impedance reaches 1MΩ/25pF, with a maximum input voltage of 300V (DC+AC peak). It is equipped with a 2Vp-p/1kHz calibration signal for easy probe calibration. Adopting Japanese electronic encoding switches and sealed attenuation switches, it is lightweight and durable, meeting high-precision signal measurement needs in laboratories, production lines, and other scenarios, with simple operation and stable reliability.
| Vertical System | Sensitivity | 5mV~5V/DIV, 10 steps in 1-2-5 sequence |
| Sensitivity and accuracy | ≤3%; 1/2.5 or smaller than the panel indicating scale | |
| Frequency range | DC~20MHz | |
| AC coupling | <10Hz (100kHz 8DIV frequency response:-3dB) | |
| Rise time | Approx. 17.5nS | |
| Input resistance | Approx. 1MΩ/25pF | |
| DC balance movement | 5mV~5V/DIV: ±0.5DIV | |
| Linear | Amplitude change <±0.1DIV (When waveform moves vertically in center) | |
| Vertical mode | CH1; CH2; DUAL: CH1 and CH2 display simultaneously (Speed alternate/intermittent) | |
| ADD | CH1 and CH2 do algebraic addition | |
| Intermittent repetition frequency | Approx. 250kHz | |
| Input coupling | AC, GND, DC | |
| Maximum input voltage | 300V peak (AC: Freq.≤1kHz) | |
| Common mode rejection ratio | >50:1 at 50kHz sine wave | |
| Insulation between 2 channels | >1000:1 (50kHz); >30:1 (15MHz); >30:1 (35MHz); >30:1 (45MHz) | |
| Horizontal System | Sweep time | 0.2μs~0.5s/DIV, 20 steps in 1-2-5 sequence |
| Accuracy | ±3% | |
| Sweeping magnification | 10 times | |
| X10MAG sweep time accuracy | ±5% (20nSec~50nSec not calibrated) | |
| Trigger System | Trigger source | CH1, CH2, LINE, EXT |
| Trigger Coupling | AC: 20Hz to full bandwidth | |
| Trigger slope | +/- | |
| Sensitivity | 20Hz~2MHz: 1DIV; TRIG-ALT: 2DIV; EXT: 200mV (2MHz~20MHz: 1.5DIV; TRIG-ALT: 3DIV; EXT: 800mV) | |
| TV | Sync pulse > 1DIV (EXT:1V) | |
| Trigger mode | AUTO, NORM | |
| X-Y Mode | Sensitivity | same as vertical axis |
| Frequency range | DC~500kHz | |
| X-Y phase error | ≤3° (DC~50kHz) | |
| Calibration Signal | Waveform/Freq | Square wave, Approx. 1kHz (Duty cycle <48:52) |
| Output voltage | 2Vp-p ±2% | |
| CRT Oscilloscope Tube | Model/Acceleration voltage | 6 inch rectangular internal graticule / Approx. 2kV |
| Characteristics | Power source | AC 220V±10% (standard) or AC 110V/220V±10% 50Hz/60Hz, 35VA max |
| Dimension/Weight | 455(W) x 150(H) x 310(D) mm / Approx. 8kg |